• Object ID: 00000018WIA30E4B770GYZ
  • Topic ID: id_40026532 Version: 1.4
  • Date: Mar 2, 2022 4:46:27 PM

Navigator: considerations and troubleshooting tips

Use the Navigator option to obtain a navigated, free-breathing, coronary artery or body acquisition. There are two Navigator modes:

  • Navigator Gating
  • Navigator Triggering

Navigator gating

A Navigator gating scan time is almost triple the scan time of a breath-hold scan.

Navigator gating is an alternate acquisition method for the navigator and imaging signal. Note that the image signal is only stored when the navigator signal is within a predefined acceptable range. Navigator gating is compatible with the following PSDs:

Figure 1. Coronary artery scans
ModePSD familyPulseImaging options
3DGradient EchoFast SPGRNavigator, Cardiac Gating/Triggering
3DGradient EchoFast GRENavigator, Cardiac Gating/Triggering
3DGradient EchoFIESTANavigator, Cardiac Gating/Triggering
Figure 2. Body scans
ModePSD familyPulseImaging options
3DGradient EchoLAVANavigator
3DGradient EchoFast GRE or Fast SPGRNavigator Imaging Option and Dual Echo (# of TEs per Scan = 2)
Figure 3. Body scan one-click application
ModeOne-click ApplicationImaging options
3DLAVA-FlexNavigator
3DDISCONavigator

Navigator triggering

Navigator triggering is a continuous acquisition of the navigator and imaging signal only when the detected position of the navigator signal is within a predefined acceptable range. Navigator triggering is compatible with the following PSDs:

Figure 4. Body scans
ModePSD familyPulseImaging option/notes
2DEcho Planar ImagingDW EPINavigator
2DFast Spin EchoFRFSE, FSE-IR, FSE, SSFSE, SSFSE-IRNavigator
3DFast Spin EchoFRFSE-XLNavigator, MRCP
3DFast Spin EchoCubeNavigator
3DVascularInhance IFIR, only available on 1.5T systemsNavigator

Navigator troubleshooting tips

Navigator signal enhancement may reduce T1 contrast if a low flip angle (approximately 10 degrees) is used.

Figure 5. Troubleshooting tips for examples of Navigator patterns
Navigator waveform patternCauseAction
Drift down

Patient specificWait 3-4 cycles and move the threshold up slightly, if needed.
Drift up

Patient specificMove threshold up slightly at once (avoids accepting mid respiratory points).
Apnea

Patient specificNo action.
Abnormally small excursion

Patient specificReduce Navigator acceptance window.
Acceptable poor signal

The low navigator signal is due to inhomogeneity effects at the dome of the liver.There is no need to change any Navigator parameters since the green waveform correctly reflects the respiratory status.
Low SNR

Saturation effect by imaging excitation.There are three options to correct the saturation effect:
  • From the Graphic Rx toolbar, increase the tracker thickness.

For LAVA, LAVA-FLEX, 3D Dual Echo, there are two additional options:

  • From the Navigator tab, click Navigator Signal Enhancement option.
  • Reduce the flip angle.
Diaphragm is far from center of grayscale

Tracker is located too far away from diaphragm.From the Graphic Rx viewport, center the tracker line between the diaphragm and the lung. For details, see:
Undesirable signal

Interference from any of the following signals:
  • Heart
  • Lung inflammation
  • Subcutaneous fat
  • Visceral fat
  • Vessels