• Object ID: 00000018WIA3024A870GYZ
  • Topic ID: id_40025610 Version: 1.4
  • Date: Aug 28, 2022 6:51:06 PM

SAR Optimization

Use SAR Optimization to reduce scan time for neuro and C-spine exams, when protocols are SAR limited. It is compatible with all Brain and C-spine coils, Body coil transmitting with the chin as landmark, and the following PSDs:

There are two options:

  • 0 = Off
  • 1 = On, with the following potential considerations:
    • reduced SAR
    • increased image blurriness
    • decreased SNR
    • decreased spatial saturation
    • increased flow artifact

With Spin Echo and T1 MEMP PSDs, there are three options:

  • 0 = Off
  • 1 = Mid
    • reduced SAR
  • 2 = High
    • reduced SAR more than Mid(=1)
    • increased T1 contrast
    • decreased SNR
    • increased CSF flow artifact