- 00000018WIA3047A870GYZ
- id_400221771.4
- Aug 28, 2022 8:23:59 PM
Spatial SAT Type
Use the Spatial SAT Type User CV with the 2D FSE family of scans (FSE, FRFSE, PROPELLER, etc.) to vary the amount of tissue suppression for spatial SAT pulses.
There are three options for 2D FSE PSDs:
- 0 = Light
- 1 = Medium
- 2 = Strong
Considerations
- Spatial SAT 1 or 2 may reduce the number of available slices per TR. Consider selecting Auto TR from the TR menu on the Scan Parameters screen to optimize the TR for a single acquisition.
- Spatial SAT 1 or 2 may result in reduced signal suppression in areas where the SAT pulses overlap. Therefore, is it recommended that you don't overlap SAT pulses.
Figure 1. Spatial SAT Type applied to 2D FRFSE
Note: This example shows Spatial SAT applied anterior and posterior to the prostate.Table 1. Image legend Item Description 1 Spatial SAT Type Light (0) 2 Spatial SAT Type Medium (1) 3 Spatial SAT Type Strong (2)