• Object ID: 00000018WIA30B86970GYZ
  • Topic ID: id_40021721 Version: 1.2
  • Date: Mar 17, 2022 2:44:31 PM

Flow Quantification Optimization

Use Flow Quantification Optimization to reduce the phase error in quantitative flow measurement for Fast 2D Phase Contrast when used with Flow Analysis.

  • Enter 1 to turn on (0 = off).
  • When Flow Quantification Optimization is on, the TE is set to minimum and cannot be changed.
  • If the Flow Recon Type = Phase Diff, Flow Direction = In slice, and Flow Analysis = On and Flow Quantification Optimization is on, the flow encoding gradients are optimized and the TE is lengthened by approximately 1 ms.