• Object ID: 00000018WIA30750970GYZ
  • Topic ID: id_40022053 Version: 1.9
  • Date: Aug 17, 2022 5:48:52 PM

MUSE

About this task

Use these steps to acquire high resolution, multi-shot DWI/DTI scans: MUSE application. It acquires a segmented scan in the phase encoding direction, which mitigates shot-to-shot motion-induced phase errors inherent in multi-shot diffusion and diffusion tensor scans. The result is images with reduced blurring and susceptibility artifacts. Note that it does not correct voluntary patient motion artifacts.

Figure 1. MUSE de-selected and selected
Table 1. Image legend
NumberDescription
1MUSE de-selected. Note the blurring and susceptibility artifacts.
2MUSE selected from the PSD/Application screen.

Step-by-step instructions

  1. Start a scan session.
    MUSE is compatible with all coils with 8 channels or more.
  2. From the Workflow Manager, click Add Task > Add Sequence.
  3. From the Protocol screen, select a MUSE protocol from your site or GE library.
    • At a minimum, include a localizer and MUSE series.
    • MUSE automatically selects ASSET and thus a calibration scan must be acquired. From the Details screen > Calibration in Prescan menu, select On. For details, see Calibration scan procedure.
    • If you cannot locate a MUSE protocol in either library, then from the Protocol screen:
      1. From the GE protocol library, click the Template tab.
      2. Click the Applications folder.
      3. From the Applications list, click MUSE and add it to the Multi-Protocol Basket.
      4. Click Accept.
  4. Acquire a 3-Plane localizer.
  5. From the Workflow Manager, select the MUSE series and click Setup.
    1. Make scan parameter adjustments, as needed.
      • From the PSD/Applications/Imaging Options screen, select any compatible Imaging Options.
      • From the details screen, select a matrix values. MUSE allows larger matrix sizes in comparison to conventional single shot DWI or DTI scans.
      • From the Details tab, select Number of Shots. The minimum number of shots is 2.
      • From the Acceleration tab, select a Phase value.
    2. Click the Advanced tab and make adjustments as needed.
    3. Graphically deposit and position the slice.
  6. If desired, see Add post-process task. There are multiple compatible post-process tasks if Diffusion Direction is selected from the Diffusion tab is set to All.
  7. When the prescription is finished, click Save Rx > Scan.

Results

Consideration this information when acquiring a MUSE scan.
  • It is not compatible with EPI FLAIR.
  • It is not compatible with PROPELLER DWI.
  • Image reconstruction time is slightly longer in comparison to a DWI or DTI scan.
  • Image annotation:
    • epi2muse
  • When all of the anatomy is not included in the phase FOV, a sharp line may appear across the images. If this occurs, repeat the scan and use a larger FOV to remove the artifact.
    Figure 2. Scan acquired with no air around Phase FOV with end result of artifact

    Scan acquired with air-to-air in the phase FOV with MUSE Diffusion. The addition of anterior and posterior saturation bands help to reduce motion and artifacts from fat.

    Figure 3. Scan with air-to-air space around FOV
  • The inherent susceptibility-induced geometric distortions in EPI-based DWI and DTI scans can be reduced by applying ASSET Imaging Option. ASSET requires a calibration scan and a mismatch between the calibration scan (RF sensitivity profile) and the EPI-based DWI/DTI scan due to susceptibility effects and motion might result in aliasing of bright structures such as eyeballs in T2 (b=0) and/or lower b-value images [1], see the following image. In the case of MUSE, potential shot-to-shot mismatch due to motion/inherent phase errors coupled with higher resolution images might exacerbate the problem such that aliasing artifacts from bright structures (e.g eyeballs) could be seen more frequently [2]. It is also important to note that derived images such as ADC and FA maps will also show this artifact if it is visible on lower b-value and/or T2 (b=0) images.

    SAT band placement over the eyeballs was reported in literature as an effective way to mitigate eyeball aliasing [1]. Care should be taken to avoid inadvertent suppression of signals from the region of interest.

    Figure 4. Eyeball artifact
    Table 2. Image legend
    NumberDescription
    1Example of eyeball ghosting in a single-shot EPI scan.
    2Example of eyeball ghosting in a multi-shot-EPI scan.
    3SAT band placement over eyeballs mitigates the eyeball artifact.
    4Mitigated the eyeball artifact from SAT band.

    [1] Chou, M-C., et al. "Pseudolesions arising from unfolding artifacts in diffusion imaging with use of parallel acquisition: origin and remedies." American journal of neuroradiology 28.6 (2007): 1099-1101.

    [2] Chang, Hing‐Chiu, Shayan Guhaniyogi, and Nan‐kuei Chen. "Interleaved diffusion‐weighted improved by adaptive partial‐Fourier and multiband multiplexed sensitivity‐encoding reconstruction." Magnetic resonance in medicine 73.5 (2015): 1872-1884.