Multi-echo FGRE/FSPGR scan
Gradient Echo pulse sequences are sensitive to magnetic susceptibility effects. When anatomical regions that have increased ferromagnetic matter (such as iron) are scanned, the result is a dark or low signal intensity in the iron containing region. As TE increases, the SNR decreases significantly.
It is possible to quantify susceptibility effects by measuring T2* relaxation time via at least two sets of images that have different TE times. 2D Multi-Echo FGRE/FSPGR enables allows selection of 3-16 echoes. These images can be post-processed in READY View to measure the relaxation time variants using READY View: R2Star workflow.
Use Multi-echo FGRE/FSPGR:
- to acquire images to analyze iron-load for blood-transfused patients
- in the clinical management of patients with iron overload diseases such as Thalassemia
- in liver and myocardium examinations
Consider this information when modifying Multi-echo FGRE/FSPGR scan parameters for images to be post-processed in R2Star READY View application. For specific scan parameter values, select a protocol from your GE or Site library.
| Warning | |
|---|---|
| CAUTION | |
|---|---|
- Scan selections: 2D Mode, Gradient Echo family, Multi-echo Fast GRE or Multi-echo Fast SPGR pulse.
Scan parameters
- # of echoes: The number of echoes range is 3-16. Typically select a minimum of 4 or 5.
- SAT: To minimize Fat/Water in-phase/out-of-phase effect, select Chemical Saturation for liver scans.
- TE: It is automatically calculated.
- Select Minimum TE for liver scans in order to get a short TE time for the first echo.
Imaging Options
The following imaging options are available with Multi-Echo FGRE/FSPGR
- ASSET
- ARC with ARC compatible coils
- Cardiac Gating/Triggering
- Extended Dynamic Range
- Flow Compensation
- Square Pixel
- Respiratory Gating/Triggering
- ZIP 512
User CVs
Click the Advance tab to view the available User CVs. The CVs may vary based on the field strength and selected scan and imaging parameters.
Post-process tasks
There are multiple compatible post-process tasks. For details, see Add post-process task.
- Image Enhancement Filters
- Parameters (Lower Threshold, Confidence level, Kernel size, Fitting algorithm) and Functional maps R2* and T2*). For details, see READY View R2Star Optional procedures
CAUTION