Prescribe and scan a pulse sequence
About this task
- The PURE feature changes the pixel intensity in the image to provide a more uniform image. For example, when cardiac protocols are used in conjunction with PURE for dynamic MRI time-course imaging, delayed enhancement, StarMap and phase contrast imaging, a modification to the AW applications of ReportCARD*, Flow Analysis, StarMap or AngioCARD difference threshold may be required. We recommend disabling PURE for all phase contrast acquisition. It is recommended to either disable PURE for dynamic time-course, delayed enhancement, StarMap studies or apply the PURE feature uniformly across all the series.
- The output of post processing can be modified with any change to the protocol. For optimal image quality, keep consistency with protocols and use the same type of filter or apply PURE for both pre and post contrast imaging.
- If using the ReportCARD, Flow Analysis, StarMap or AngioCARD post processing software, refer to the operator manual regarding any change to your imaging protocol and the use of PURE or other image filtering techniques.
Use these steps to prescribe and scan a pulse sequence.