AutoStart Modality Worklist

About this task

Use these steps to automatically start the first series of an exam to reduce the time the patient spends on the table in the scan room.

Step-by-step instructions

  1. Select the patient from the Modality Worklist.
    • If the patient is not in the Modality Worklist, follow these steps:
      1. Click the New icon.
      2. Enter patient demographic information. At a minimum, the patient’s ID and correct weight must be entered.
      3. Click Show All Protocols/Modify Protocol and select a protocol that has AutoStart. Note that there is no indication that a protocol has AutoStart tuned on, unless you have added it to the protocol name.
      4. Click Start Exam.
  2. Select the Exam dB/dt and SAR Limits and click Accept.
  3. If AutoStart has been turned on in the protocol, AutoStart should be selected.
    • If AutoStart is turned on in the protocol, the prompt Do you want to keep AutoStart enabled? displays. Click Yes to continue.
    • If AutoStart was not turned on in the protocol, select AutoStart.
    • Only the first series in an exam can have AutoStart applied.
  4. Complete all fields for the first series in the protocol.
    • Make sure the coil on the table matches the coil in the protocol.
    • If any required fields are not completed, AutoStart will not begin the first series.