Prescan
Prescan optimizes the system’s transmit and receive values for each patient and area of interest. Every patient is a unique collection of protons and requires a unique combination of RF and power to image.
There are multiple prescan options, which are selected from the Scan menu:
- Auto Prescan
- Manual Prescan
- Spectroscopy Prescan
Figure 1. Scan menu 
The primary reasons for prescan are:
- to include a calibration scan for tasks that have a PURE filter, an ASSET imaging option, or both.
- to adjust the system settings (center frequency, transmit gain, receive gain etc.) for optimum image quality
- to optimize auto shim values (when selected) for homogeneity improvement of the magnetic field with the patient in the bore
- to allow the manual optimization of chemical saturation