• Object ID: 00000018WIA3062E870GYZ
  • Topic ID: id_40025259 Version: 1.4
  • Date: May 20, 2022 9:46:49 AM

Vector gating setup

Step-by-step instructions

  1. Connect the electrodes to the black and white lead pairs that are marked by a color bead located near the connector. The leads are color coded, but also have a color center stud on the wire near the connector. The leads should be connected as orthogonal pairs as shown in the image below. They can be in any direction as long as the pairs are orthogonal.
    • white lead: white stud
    • red lead: white stud
    • green lead: black stud
    • black lead: black stud
      Figure 1. Vector electrode placement: arrows represent Vector 1 or 2, as long as the pairs are orthogonal.
  2. In the footer area, click the Gating, Fan and Light controls icon to open the Gating Control screen.
  3. Click Cardiac Gating.
  4. For wired gating, click Independent Vector Gating.
    Both waveforms are displayed on the Waveform screen.
  5. Confirm that the waveforms are acceptable. Adjust electrode placement, if necessary.
  6. Advance the patient to the scan location.
  7. Select Gating Reset to clear the triggering history and relearn the patient’s waveforms.
    Do this before scanning when the patient is at the scan position and the waveforms are free of interferences.