Cardiac and respiratory gating/triggering

The following pulse sequences allow the use of Combined Cardiac and Respiratory Gating/Triggering: 3D Fast Gradient Echo, Fast Card, and FastCINE sequences.

If a valid ECG trigger is detected within the respiratory data acquisition window, scanning begins. When the end of the data acquisition window is reached, scanning stops. RF application is continuous, being applied at the TR time throughout the entire respiratory cycle, even when the data acquisition window is closed.

Figure 1. Combined Respiratory and Cardiac Gating/Triggering
Table 1. Respiratory/Cardiac image legend
NumberDescription
1One respiratory cycle from 0-100%
2Trigger Point of 30%
3Time frame when cardiac gated slices are acquired
4Trigger Window
5Time frame when cardiac gated slices are acquired
6Slices