• Object ID: 00000018WIA30ECE770GYZ
  • Topic ID: id_40021763 Version: 1.4
  • Date: May 5, 2022 3:57:36 PM

Cardiac introduction

Cardiac Gating/Triggering allows the same data to be collected at the same points of the cardiac cycle for each repetition of the pulse sequence allowing motion artifacts to be reduced, or sometimes eliminated. The benefit is that the motion is always the same and is essentially “frozen,” allowing motion artifacts to be reduced or, sometimes, eliminated. Gating and triggering of MR images can be performed with ECG or Peripheral gating techniques.

Table 1. Cardiac applications
Scan planeCardiac application
Short AxisAssesses two chambers, either both atria or both ventricles.
2-Chamber Long AxisVisualizes left atrium mitral valve and left ventricle.
4-Chamber Long AxisAssesses septal wall and visualizes the four chambers, mitral, and tricuspid valves in a single plane.
3-Chamber View

(left ventricular out-flow tract)

Assesses left atrium, mitral valve, left ventricle, aortic valve, and aortic root in one plane.

Cardiac Triggering

Cardiac triggering is a technique that turns RF application and image acquisition on and off based on a trigger detected within the cardiac or respiratory cycle. The Trigger Window is set as a percentage of the cardiac (or respiratory) cycle. Trigger Delay is usually set to minimum to minimize acquisition delays, unless the application requires it. Triggering synchronizes data acquisition with the waveform cycle, and all image data is acquired at the same phase in the cycle for each slice prescribed.

Spin Echo, Fast Spin Echo, Fast GRE-ET, and multi-planar 2D GRE/SPGR sequences use Cardiac Triggering.

Figure 1. Cardiac Triggering
Table 2. Image legend
NumberDescription
1Trigger Window
2RR Interval
3Trigger Delay
4Slices

Cardiac Gating

Cardiac gating monitors the cardiac or respiratory cycle, but does not use a trigger to initiate RF application. RF is applied throughout the cycle at the defined TR, and data is acquired when the desired physiological event occurs.

Fast Card, FastCINE, 3D Fast GRE/SPGR, 3D Fast TOF (GRE/SPGR), FIESTA, and Fast 2D PC use Cardiac Gating.

Figure 2. Cardiac Gating: 1 = Trigger Window, 2 = Uniform TR
Table 3. Image legend
NumberDescription
1Trigger Window
2Uniform TR

Cardiac and respiratory gating/triggering

The following pulse sequences allow the use of Combined Cardiac and Respiratory Gating/Triggering: 3D Fast Gradient Echo, Fast Card, and FastCINE sequences.

If a valid ECG trigger is detected within the respiratory data acquisition window, scanning begins. When the end of the data acquisition window is reached, scanning stops. RF application is continuous, being applied at the TR time throughout the entire respiratory cycle, even when the data acquisition window is closed.

Figure 3. Combined Respiratory and Cardiac Gating/Triggering
Table 4. Image legend
NumberDescription
1One respiratory cycle from 0-100%
2Trigger Point of 30%
3Time frame when cardiac gated slices are acquired
4Trigger Window
5Time frame when cardiac gated slices are acquired
6Slices