RF Inhomogeneity artifact
Failure of an RF coil can cause intensity variation across an image. This often indicates the failure of a coil element or the presence of ferromagnetic material in the imaged object.


Check with the patient to make sure that nothing on them or in them is causing the artifact. If you suspect a faulty coil element, check each element using manual prescan. If you isolate a faulty element discontinue using the coil and consult with your service engineer.