TOF tab

The Vascular TOF tab only appears if the protocol is a TOF pulse sequence. Only the vascular selections that are compatible with the active protocol display on the Vascular tab.

Figure 1. Scan Parameters: Vascular 3D TOF tab
ItemDescription
Projection ImagesChoose the desired number of Projection Images or leave this blank and create projection images in IVI.
CollapseCollapse reconstructs a collapsed view in the plane of acquisition.
Ramp PulseSee below.

Ramp Pulse

Ramp Pulse uses a variable flip angle over the imaging volume to reduce saturation of incoming flow. Ramped flip angle excitation combines the benefits of custom-designed, minimum-phase RF pulses with the flow selectivity and increased background suppression of larger flip angles.

  • A ramped flip angle excitation pulse with slope 2 has a flip angle that doubles from entry to exit slice. The pulse center flip angle is prescribed; the entry and exit slice flip angles are 2/3 and 4/3 of the center flip angle, respectively. Spatially-varying excitation flip angles prevents saturation of slowly-flowing blood at the entry (low flip angle) part of the slab and provides suppression of venous flow with the large flip angle (exit) portion of the slab. These effects allow improved visualization of slow, in-plane, and tortuous flow, and eliminate the need for a spatial presaturation pulse to cancel venous blood signal.
    Figure 2. Example using a 30° flip angle ramped RF pulse
    Table 1. Ramp Pulse image legend
    ItemDescription
    140° RF pulse [30 + (1/3 x 30) = 40]
    220° RF pulse [30 - (1/3/ x 30) = 20]
    3Slice number 60 with a 40° flip angle
    4Slice number 30 with a 30° flip angle
    5Slice number 1 with a 20° flip angle
  • Do not use Ramp Pulse when the flip angle in your protocol is between 46 and 90. In general, choose flip angles in the 20 to 45° range with Ramp Pulses.
  • 3D TOF is less sensitive to flow, which may become saturated as it passes through the imaging volume. The use of Ramp Pulses can offset some of this saturation.
  • Ramp Pulses produce images with a variation in background suppression from entry to exit slice and flow speed selectivity as a function of location within the slab.
  • Slab boundary artifacts are more prominent with Ramp Pulses.
  • Ramp Pulses and Magnetization Transfer (in head) increase small intracranial vessel conspicuity when combined with a high-resolution acquisition.
  • Ramp Pulse offers the following image quality improvements:
    • Improved slab profile by reducing wrap and improving image quality away from the center of the slab
    • Reduced saturation of blood flow increasing the visibility of arteries
    • Reduced sensitivity to venous signal
    • Shortened minimum TE
  • The table below can help you determine the ramp pulse direction, depending on the direction of the blood flow and whether you need to see bright blood or dark blood.
    Table 2. Ramp Pulse direction
    Ramped Pulse directionBlood flow directionSignal intensity
    I-SI-SBright
    I-SS-IDark
    S-II-SDark
    S-IS-IBright
  • Select the Ramp Pulse flow direction based on the scan plane selected. Refer to the following table for guidance.
    Table 3. Ramp Pulse direction based on scan plane selection
    Scan plane Ramped Pulse directionBlood flow direction
    AxialI to S Flow and S to I Flow
    Head Coil AxialI to S Flow is defaulted on
    SagittalI to S Flow and S to I Flow
    CoronalP to A Flow and A to P Flow