Required characteristics and configuration for network support of MR scanner

The network must meet the specific requirements above for all traffic flows associated with the subset of features, use cases and workflow required by the responsible organization’s users.

In addition, the network must be “flat” (limited to a single IP broadcast domain).

  • Connection of the MR System to a network that includes other equipment could result in the loss of supported DICOM functionality if network does not meet the defined specifications.
  • The hospital should ensure proper network configuration.
  • Subsequent changes to the network could introduce new risks and require additional analysis.
  • Changes to the network includes:
    • Changes in the network configuration.
    • Connection of additional items to the network.
    • Disconnecting items from the network.
    • Update of equipment connected to the network.
    • upgrade of equipment connected to the network.