In-room display gating screen

Figure 1. Example of: Gating screen
Table 1. Gating screen selections
NumberDescription
1ECG icon : select to toggle the display between all four waveforms and the cardiac-only waveform.

Respiratory gated icon : select to toggle the display between all four waveforms and the respiratory-only waveform.

Peripheral gated icon : select to toggle the display between all four waveforms and the peripheral gated only waveform.

2Waveform display: ECG gating, peripheral gating, and respiratory.
3Click to select gating lead: VCGI, VCG-II, VCG I + II, PG, and RESP.
4Click to invert the waveform.
5Settings icon : select to toggle the Settings control panel on/off.

Reset icon : select to reset the gating system.

Settings control panel

The options on the Settings control panel can also be accessed from the Gating Control screen. The Settings control panel allows you to do the following:

  • Select the best trigger lead.
  • Invert the ECG waveform
  • Select the cardiac gating lead type: VCG, Standard or third party.
  • Apply a noise filter for Standard gating leads.
Figure 2. Example of: Settings control panel