DAQA messages

The following table displays the content of DAQA messages.

NumberMessage
1.A different coil is connected! Please re-select the coil and plane before hitting [Start].Coil not valid.
2.A different coil is connected! This coil is also not supported by Daily QA Tool. Please plug in a valid supported coil before hitting [Start].Coil not valid.
3.Aberration Geometric Distortion
4.Abort
5.ATP execution has hung, please check atp process or the scanner hardware
6.but was called with incorrect input arguments., ...
7.Center not within the object. This algorithm works best for convex objects. Very likely that your results are in error
8.Center Frequency
9.Close
10.Coil configuration notice
11.Daily Automated Quality Assurance
12.Daily QA Tool aborted on date_time
13.Do you really want to abort the current scan?','Abort?
14.ERROR
15.Error building SVAT file for manual prescan.
16.Error building SVAT file to load the protocol.
17.Error editing the protocol for COIL
18.Error editing the protocol for FOV
19.Error editing the protocol for GRADMODE
20.Error editing the protocol for PLANE
21.Error editing the protocol for SWAPPF
22.Error executing the SVAT script to load the protocol.
23.Error executing the SVAT script to run Auto-Prescan.
24.Error executing the SVAT script to run first image scan.
25.Error executing the SVAT script to run Manual-Prescan.
26.Error executing the SVAT script to run second image scan.
27.Error from get_coilid function.
28.Error from read_coil_id_list function.
29.Error with abort_svat function
30.ERROR with APS_EVENT svat command
31.ERROR with DOWNLOAD svat command
32.ERROR with DOWNLOAD svat command, scanner busy
33.ERROR with IPG_ADVANCE_TOSC svat command
34.ERROR with LOADPROTOCOL svat command
35.ERROR with MODIFY_CV svat command
36.ERROR with MPS_SCAN_TR svat command
37.ERROR with NEW_EXAM svat command
38.ERROR with PROTOCOL_DIR svat command
39.ERROR with PROTOCOL_MODE svat command
40.ERROR with PSC_UPDATE_VAL svat command
41.ERROR with RECON_STOPPED svat command
42.ERROR with RESET_SCAN svat command
43.Error with reset_svat function
44.ERROR with SCAN_EVENT svat command
45.ERROR with START_LOOP_EVENT svat command
46.ERROR with STOP_LOOP_EVENT svat command
47.Error with table_wait_time function
48.ERROR with VIEW_EDIT svat command
49.Error! Cannot find the phantom!
50.Error! Two test images do not have the same size!
51.Exit
52.Ghosting Level
53.Ghosting Level & Geometric Accuracy
54.Images: too Few Inputs
55.Images: too Many Inputs
56.Incomplete Rx, please check coil and landmark
57.Max
58.Maximum Geometric Distortion
59.Mean
60.Min
61.No
62.No P files found!!!
63.No valid landmark
64.Noise
65.
Note: Phantom placement and coil landmarking are critical for repeatable results. Verify coil and phantom are properly placed and landmarked at correct location. Also verify there are no large air bubbles in the phantom. Do you wish to continue?
66.OK
67.Only system configuration is allowed for ghosting level option. Make sure you landmark on this configuration or unselect ghosting level option. SNR is measured on Axial plane
68.Please Select a DAQA history file
69.Please select the coil ! Select Coil
70.Please select the coil and scan plane!.Select Coil & Plane
71.Please wait at least 15 minutes before scanning to prevent swirling artifacts. Do you wish to continue?
72.Please wait while the DAQA initializes..
73.Protocol download failed
74.Result Files
75.Scale Geometric Distortion
76.Select coil
77.Select Scan Plane
78.Signal
79.SNR
80.Standard Deviation
81.Start
82.Success: Center within the object!!!
83.Sufficient Data is not available for trending. Atleast two runs of the DAQA tool is required.
84.Table check has hung, please check table_feedback process or the scanner hardware
85.Test Completed
86.Test Completed
87.Test Completed! Results are recorded in output file
88.Test proceeds without at least 15-minute waiting time. Test result might not be stale','Test proceeds without enough waiting time
89.The current coil is not supported by Daily QA Tool. Please plug in a valid supported coil before hitting [Satrt].Coil not valid.
90.There is not that many files listed !!!
91.There might be artifacts on the test images or you are not using homogeneous phantoms!
92.Transmit Gain
93.Trend Data
94.Trend Viewer
95.Wait for a minute after moving the table
96.Yes
97.You chose to include ghosting level and geometry accuracy in addition to SNR in the test. The test will use connected coil configuration and 17 cm sphere phantom with loader if Head Coil is used, otherwise 17 cm sphere phantom without loader. Do you want to continue?
98.You chose to include ghosting level and geometry accuracy in addition to SNR in the test. The test will use connected coil configuration and 27 cm sphere phantom with a loader. Do you want to continue?
99.You chose to use connected coil configuration and Axial plane. Please make sure that you use a 17 cm sphere phantom with loader if Head Coil is used, otherwise 17 cm sphere phantom without loader. Do you want to continue?