Emergency stop recovery procedure

About this task

Use these steps to recover normal system operation after an emergency-stop has occurred. For emergency stop details, see System emergency logon.

Step-by-step instructions

  1. Go into the equipment room and locate the ISC cabinet.
    • The equipment room is typically only used by service engineers to service the equipment. Use caution when inside the equipment room.
  2. From the ISC cabinet, which contains the system PDU and Gradient PDU, complete the following.
    1. Open the small access door to view the Main Breaker and EMO button.
    2. Press the green EMO Reset button.
      Figure 1. ISC cabinet
      Table 1. Image legend
      NumberDescription
      1ISC cabinet
      2Main Breaker in On position, fully vertical.
      3EMO Reset button.
  3. Return to the operator console and complete a TPS reset. For details, see TPS reset.
  4. After the TPS reset is finished, the emergency-stop condition should be recovered. If the emergency-stop condition remains after the TPS reset is finished, call your service engineer.