T2 Map scan protocol

  • When using Multi-echo FGRE/FSPGR for tissues with short T2* relaxation times, the echo spacing and first echo should be as small as possible (around 1 ms for first echo time) to ensure confidence in the fit of the R2* value. The best way to achieve this is by using a high band width, low matrix, and not using User CV-16, positive gradient polarity.

Consider these tradeoffs as you change scan parameters for images that are post processed with the T2 Map protocol.

Table 1. T2 Map protocol example
Scan parameters1.5T valueConsiderations
TR1000 msTo shorten scan time, it is possible to reduce TR much more than what would be normally needed for T2 imaging. This is because the T1 relaxation time of cartilage is quite short (in the range of 200 to 400 ms), and thus there is almost no T1 saturation in cartilage when reducing TR.

Reduce TR below the suggested values, decreases the overall SNR and also introduces some T1 saturation effect that could generate additional slice cross talk effects that affect T2 relaxation.

TEDetermines the number of images acquired at each locationTE is the primary scan parameter that controls T2 image contrast. It is not a selectable parameter with the T2 Map application.
Bandwidth32
Field of View16The value does not change for field strength.
Slice thickness3
Slice spacing0If slice spacing is less than 20% of the slice thickness and if the TR prescribed can accommodate the prescribed slices in a single acquisition, the system will automatically change the prescription to two acquisitions in order to reduce the cross talk effects.
Frequency256
Phase256Consider a 192 phase matrix to reduce the total scan time below 7-8 minutes.
Scan planeAxialPatellofemoral Joint
Scan planeSagittal or CoronalFemoral Condyle and Tibial Plateau
Imaging optionsASSET or ARCA Phase factor > 1.0 on the Acceleration tab can be used in order to reduce scan time. Any larger acceleration factor might rapidly degrade SNR and therefore the T2 Map image quality.

If available, use AIR Recon DL to increase SNR of this sequence, especially when phase acceleration factor >= 2.0.