Secondary magnet rundown

About this task

Use these steps with the vacuum break tool to break the magnet vacuum for a magnet rundown.
Important: This secondary quench method should be attempted only after other approved methods of rundown have been attempted and in the event of a life threatening condition. Performing this procedure will take the magnet out of service for at least one month.
  1. If possible, call your qualified service representative.
  2. Locate the vacuum break tool (46-260852G3).
    Note: This tool is for emergency only and should be stored in a safe and easily accessible location. Ensure that all operators are informed of this tool's location.
    Figure 1. Vacuum break tool
  3. Ensure all room ventilation systems are on and the scan magnet room door is propped open.
  4. Locate the vacuum port access cover on your magnet cover.
    Figure 2. Vacuum port access on the magnet cover
  5. Remove the vacuum port access cover from the magnet cover to expose the vacuum port.
    • You should be able to remove the access cover with your finger or non-magnetic tool.
      Figure 3. Example of an access cover off and the vacuum port exposed
  6. Remove the protective dust cover from the vacuum port.
    • The magnet cover in this image is shown removed for illustration purposes only.
      Figure 4. Dust cover removed
  7. Screw the handle of the vacuum break tool into the plug of the vacuum port by turning it clockwise.
    • At least three full turns are required to ensure thread engagement.
    • If you are unable to turn the handle, either call a GE service representative for further assistance or remove system enclosure cover.
      Figure 5. Vacuum break tool screwed into vacuum port plug
  8. Lift the vacuum break tool handle until it is parallel to the floor. Leave the handle in this position.
    Figure 6. Vacuum break tool parallel with floor
    • There will be an inrush of air into the magnet and the magnet will quench in about 30 seconds.