Planned maintenance service schedules

Specific customer requirements and/or your site environment may necessitate more or less frequent intervals for planned maintenance (PM) service. An agreement to perform PMs less frequently than these recommendations can be made with the understanding that a reduction of system performance may result.

To further enhance productivity, the PM procedures have been divided into four types:

  • Type 1 – Safety and Regulatory
  • Type 2 – Image Quality
  • Type 3 – Other (System not available for patient scans)
  • Type 4 – Other (System available for patient scans)

The PM matrices list all the PM procedures and the frequency they should be completed, according to schedules listed below. The schedules indicate the procedures that are performed during each visit. They also show the type (1 - 4) for each procedure. The services should be completed at the indicated intervals and should be performed only by qualified service personnel.

The PM service schedule lists all the PM procedures and the frequency they should be completed according to schedules A -D. The checks are on a sequential yearly schedule, for example A is the first three months, B is the second three months, etc.

Table 1. Four/year PM schedule
ScheduleTimeframe
A0 to 3 months
B4 to 6 months
C7 to 9 months
D10 -12 months
Table 2. 1.5T SIGNA Voyager service schedule
SubsystemPM TaskTest purposeABCD
Magnet Room
Magnet RoomOxygen Monitor Operation (Option)SafetyX
Magnet RoomPAC Leakage Curretn TestSafetyX
Magnet RoomChecking the ECG leadsSafetyXX
Magnet RoomPneumatic Patient Alert SystemSafetyXX
Magnet RoomChecking the Emergency Exhaust VentSafetyXX
Magnet RoomChecking the RF Screen Room DoorCheckX
Magnet RoomReplacing and Cleaning the Blower Box Filter (Only for Voyager)ReplaceX
Cryocooler Compressor Test
EquipmentInspect Cryocooler AdsorberSafetyXX
Integrated System Cabinet (ISC)
RF PowerUPM Function CheckSafetyX
RF PowerUPM Function Check (MNS) (Option)SafetyX
Patient Handling
Patient HandlingCradle Egress FunctionalitySafetyXX
Patient HandlingCheck the Patient TableCheckXX
Patient HandlingChecking the Cradle Home SensorCheckXX
Integrated Cooling Cabinet (ICC)
Cabinet Cooling UnitCheck Fluid Levels and Top offCheckXXXX
Facility Plumbing Unit FPU Manifold Flush/Cleaning/Check (ICC Cabinet)CheckXX
Integrated System CabinetISC Manifold Flush/Cleaning (ICC Cabinet)CheckXX
Gradient Cooling UnitGCU Coolant DeionizationReplaceX
Gradient Cooling UnitCCU Coolant DeionizationReplaceX
ICC Penetration PanelBody Coil Blower and air filterCheckXXXX
ICC Penetration PanelPatient Blower and Air FilterCheckXXXX
PDU
PDUEmergency Off & Stop Circuits/IndicatorSafetyX
Cabinet MonitorLeak Sensor Functionality CheckSafetyX
PDUPDU Power and Ground Connection Checks-Check and Tighten all PDU Power and ground connections SafetyX
PDUClean GPDU & SPDU Filter (Only for Premier)CleanX
Computer
GOCClean DustCleanXX
GOCGOC Battery Replacement CheckCheckX
Magnet
Magnet & CryogensInspect Cryogen VentCheckX
Magnet & CryogensIf equipped with Remote MRU, visually inspect cable connections between remote MRU and primary MRUCheckX
Magnet & CryogensTest GE Magnet Rundown Unit (MRU). Perform Quarterly Service Test found in the MRU Manual 5265188 or 46-318393SafetyXXXX
Magnet & CryogensTest GE Rundown Unit (MRU). Perform Yearly Service Test found in the MRU Manual 5265188 or 46-318393SafetyX
Magnet CoversCyrocooler Compressor Absorber (Check Time)ReplaceXX
Magnet CoversInspect for Discoloration the inside of the bore and the front and back of the magnetCheckX
MDP
EquipmentMDP Functional Check
Note: Refer to the MDP manual for the maintenance information.
CheckX
EquipmentMDP Power & Ground Connection and Battery Replacement Check
Note: Refer to the MDP manual for all the maintenance information.
CheckX
Coil Test
Magnet RoomP-Port Cleaning and LubricationCleanXX
Image Quality
Magnet RoomLaser Light AlignmentImage QualityXX
Magnet RoomDQA CalibrationImage QualityXX
Magnet RoomDAQA CheckImage QualityXXXX
Magnet RoomMagnet Shim checkImage QualityXXXX
Magnet RoomChecking for spike noiseImage QualityXXXX
Magnet RoomDual Drive Quadrature Calibration (Not for Voyager)Image QualityXX
Image QualityConfirm that the application software revision installed on the system is the same or at a later version of the Software revision required in the "ePM form data" tab that is located in Software matrix DOC1667089. This document is available from the Common Documentation Library. Confirm that the service pack revision installed on the system is the same or at a later version that the service pack revision required in "ePM form data" tab that is located in Software matrix DOC1667089. This document is available from the Common Documentation Library.SafetyXXXX