- 00000018WIA30800870GYZ
- id_400223971.8
- Jul 26, 2022 1:17:10 PM
Cryogen and quench concerns
With superconductive MR systems, another concern related to the static magnetic field is a quench of the cryogens. A superconductive magnet uses cryogens to super-cool the electrical conductor that creates the magnetic field. Temperatures as low as -269°C (-452°F) are achieved to create the proper environment within the magnet. A quench, which is a sudden boil-off of the entire volume of cryogenic liquid, causes a rapid loss of the static magnetic field.
Liquid cryogen hazards
Cryogens come in large vacuum containers called dewars. Liquid helium is generally used for cooling purposes, although some service procedures also require liquid nitrogen. Nitrogen dewars weigh from 400 to 500 pounds when full. Helium dewars weigh from 700 to 800 pounds. In addition to large dewars, there may be smaller helium gas cylinders present. This helium gas is used to fill the magnet to proper cryogen levels. Special considerations should be observed when dealing with cryogens.
| CAUTION | |
|---|---|
| CAUTION | |
|---|---|
Quench vent failure hazards
ID # US256628
| Warning | |
|---|---|
LHe will transform into a gaseous phase if a vacuum breach occurs. When this happens, the normal room temperature in the vacuum space will cause the liquid helium to quickly boil away and must be vented out of the building. Never install a magnet to an uncompleted vent and if a vent line has a failure, do not perform any service operations on the magnet and system.
LHe will also transform into a gaseous phase when the electrical energy stored in the magnet cryostat is released. A rapid release of wire current is defined as a magnet quench. A quench can occur when part of the superconducting coil enters the normal (resistive) state. When this happens, the electric current dissipates as heat and the liquid Helium quickly boils away and must be vented out of the building.
The very cold temperatures, dynamic forces, and internal pressures produced by this expanding gas must be accommodated by the customer supplied cryogen vent system.
Room safety is ensured when a quality and compliant cryogen vent system is connected to the GE magnet to the customer supplied cryogen vent. The cryogen vent will allow the Helium gas to flow out of the MR scan room to unoccupied spaces.
Cryogen management is a safety issue. All GE and Medical Device suppliers must follow IEC regulations.
Quenches are indicated by a loud noise, warning message, or the tilting of an image on the display screen. A quench is a hazard only if the vent fails. This would present a potential asphyxiation hazard to both the patient and personnel.
It is critical to have a well-planned method to quickly evacuate the patient and personnel from the magnet room should a quench occur.
- If a quench occurs and the vent does not fail:
- The cryogen vent rapidly emits a helium exhaust plume.
- The level of helium displayed on the meter falls.
- Prescan fails.
- The system emits no gradient sound.
- If a quench occurs and the vent fails:
- The system releases white clouds of cryogen vapor into the magnet room.
- Frost forms above the magnet.
- The level of helium displayed on the meter falls.
- The oxygen monitor emits an alarm.
