- 00000018WIA30313C80GYZ
- id_20363081.2
- Feb 1, 2022 6:31:03 PM
Troubleshooting
The recommended amplitude of the VCG trace is 1.0 mV.
- A satisfactory trace displays a green waveform.
- If the signal is less than 0.8 mV, the system displays a yellow waveform.
- If the trace is less than 0.4 mV, the system displays a red waveform.

In this example, VCGI has a trace of 1.58 mV and a satisfactory green waveform. The white tick marks indicate successful triggering.
VCGII has a trace of 0.8 mV and a yellow waveform. The absence of tick marks indicates inadequate triggering.
Note: If the yellow waveform produced tick marks, the exam may continue without any further intervention. In this example, the VCGII electrode should be evaluated. Options to explore include:
- Clean the chest again thoroughly. Be sure that all hair has been adequately removed.
- Place a new electrode.
- You may need to reposition the electrodes. Consider whether pathology, such as infarc-tion, atrial fibrillation, or anatomical variance, may cause an abnormal conduction.
Contact your clinical application specialist for additional support.
