• Object ID: 00000018WIA30334F80GYZ
  • Topic ID: id_2037699 Version: 1.3
  • Date: Mar 27, 2022 10:44:56 AM

Table emergency release

Note: that the table can be lowered and raised with foot pedals in normal conditions. However, when a manual cradle removal occurs, the foot pedals to raise and lower the table temporarily do not work until the cradle reaches the home location.

Pull the table transport emergency release lever.

Figure 1. Table transport emergency release
Table 1. Image legend
NumberDescription
1Up/down pedals
2Transport/emergency release lever
The table can be lowered and raised with foot pedals in normal conditions. However, when the manual cradle release lever is used, the foot pedals no longer work to raise and lower the table because the cradle driving mechanism is retained in the magnet.
Figure 2. Table pedals
Table 2. Image legend
NumberDescription
1Up pedal
2Down pedal
Important: The patient table of the system is permanently fixed to the magnet system. Always have a non-ferrous gurney placed outside the magnet room for emergency patient transportation.