Federal law restricts the sale, distribution, and use of this device to or on the order of a physician.
CAUTION
Do not load non-system software onto the system computer.
Warning
The MR system is not designed to provide information for clinical stereotactic use. The spatial accuracy obtainable with your MR system may not be adequate for stereotactic procedures and can vary depending on the patient, the pulse sequence used, and the system itself. It is therefore recommend that MR images not be used for stereotactic applications.
Warning
Electrically conductive stereotactic devices may lead to high localized SAR. Excessive transmit power may result from interactions between the structure and the transmit coil. In addition, improper padding between the patient and any conductor may lead to excessive localized heating.
Note: Clinical stereotactic use refers to being used in localization for surgical procedures.