Overview of operational function checks
The planned maintenance (PM) services prescribed in the PM schedules represent the current manufacturer's recommendations. Specific customer requirements and/or your site environment may necessitate more or less frequent intervals for PM service. An agreement to perform PMs less frequently than these recommendations can be made with the understanding that a reduction of system performance may result.
The PM service schedules in Planned maintenance service schedules list all the PM procedures and the frequency they should be completed by qualified service personnel. There are different schedules for each system type.
You should perform the below operator function checks listed in the table below.
| Function operational checks | Description | Service interval |
|---|---|---|
| General |
Check table emergency release. |
Monthly |
| General (Cleaning) | System clean and disinfect. | Daily/following use |
| MRU Batt and LED Check | Check the MRU battery and LED lights | Weekly |
| Patient cradle and pads | Check for cleanliness of pads and clean the inside of the cradle. | Daily |
| Coils, pads, and straps | Clean with non-abrasive cleanser. Clean coil anti-skid pads with water and mild detergent only. | Daily |
| Exhaust vent fan function | Check the emergency exhaust fan operation | Weekly |
| Exhaust vent fan and dust cleaning | Scan room exhaust clean of fan and exhaust vent | Anually |
| MDP Power and Ground (and battery) | Qualified electrician check MDP for Power and Grounding and battery for replacement (if present). Refer to the system MDP manual. | Anually |
| Patient table | Clean the inside of the cradle. Check the table for proper operation. | Daily Monthly |
| Coils and coil cables | Clean with non-abrasive cleaner. Clean coil anti-skid pads with water and mild detergent. | Daily |
| Pneumatic Patient Alert | Check for defects and damage, worn cable or exposed wires. | Daily |
| Image quality | Perform quality assurance and functional checks. | Daily or as recommended |