PROBE-P edit
About this task
Use these steps to edit a PROBE-P GRx scan.
Step-by-step instructions
- Display the desired localizer images in Graphic Rx.
- Place the cursor over the image that represents the scan plane in the protocol and click to deposit the VOI.
- For example, if the protocol is for an axial scan plane, deposit the VOI on an axial localizer image.
- To make sure the Spectroscopy localizer launches in READY View , it is best to only change the VOI location and size on the image that has the CSI grid displayed. This is the image on which you deposit the VOI.
- If you need to change the voxel thickness, change it from the Voxel Thickness text box and do not change the VOI depth on the Graphic Rx images.
- If you deposited the VOI on an axial image, then do not change the VOI size on the coronal or the sagittal.
- You can change the VOI size on the axial because this is not changing the depth of the VOI.
- Do NOT change the Center Z value.
- For Single Voxel and 2D CSI prescriptions, do not change the CSI Slice Thickness value. It does not mean anything. The Voxel Thickness value determines the depth of the VOI.
- Click and drag the VOI over the area of interest on the plane where you deposited it.
- To view the single voxel explicit location, select Start/End.
Figure 1. Start/End radio button 
- The explicit fields cannot be edited.
- To explicitly define the center of the VOI and the length in all directions, select Center/Length and edit the text fields.
Figure 2. Center/Length radio button 